dc.description.abstract A method of spectroscopic analysis of trace elements in solutions has been developed. The solution is converted to a mist and analyzed by proton induced x-ray flourescence in a helium atmosphere. Related problems in the development of proper droplet size and apparatus design are discussed in detail and in chronological order. A sensitivity of the order of 1 ppm can be achieved with the final apparatus. A qualitative discussion of the bremsstrahlung limiting the sensitivity is also included.
dc.format.extent 63 pp
dc.language.iso eng
dc.title Trace element analysis of liquid samples, by proton induced x-ray fluorescence
dc.identifier.digital RICE1680
dc.type.genre Thesis
dc.type.material Text
thesis.degree.department Physics
thesis.degree.discipline Natural Sciences
thesis.degree.grantor Rice University
thesis.degree.level Masters
thesis.degree.name Master of Arts
dc.format.digitalOrigin reformatted digital
dc.identifier.callno Thesis Phys. 1973 Boustany
dc.identifier.citation Boustany, Robert Julio. "Trace element analysis of liquid samples, by proton induced x-ray fluorescence." (1973) Master’s Thesis, Rice University. https://hdl.handle.net/1911/104054.